| Model Number |
29417 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for analysis.The hub, shaft and tip were microscopically examined.The device showed multiple kinks and bends.The device also showed stretching and separations located 24.5 and 26.5cm from the hub.The device was not completely separated the inner liner was still attached.The device could not be functionally tested in a guiding catheter due to the damage on the device.Inspection of the remainder of the device revealed no damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 07may2020.It was reported that the device could not cross lesion.The 50% stenosed target lesion was in an acute angled and moderately calcified renal artery.A 105/3.0/2.5/.021/20 renegade was selected for use.During procedure, it was noted that the device could not cross the lesion due to the gap between guidewire and microcatheter.The procedure was completed with a different device.There were no patient complications reported.However, device analysis revealed device separation.
|
| |
|