MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY PDX CYCLER ASSY (NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY PDX CYCLER ASSY (NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number 180237
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the inside of the cassette door of their cycler during fill 1 of their pd treatment.The patient reported receiving an air detected in cassette alarm during drain 0 of 6 of treatment 10 times and the patient reported that they received a m83 pump a vs.B volume alarm during fill 1.The cycler was rebooted and treatment was cancelled.The patient reported that they bypassed from drain 0 to fill 1.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was available.Upon follow up, the patient stated that they are trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed, and confirmed that they were able to complete peritoneal dialysis treatment using continuous ambulatory peritoneal dialysis (capd) in the absence of the cycler.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient has received a new cycler which is working well and is continuing peritoneal dialysis therapy with no further issues.The patient confirmed that the leak was discovered in fill 1 after cancelling treatment.The liberty pdx, integrated set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.The cycler was returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
Additional information: d10, h3 plant investigation: the actual device was returned to the manufacturer for physical evaluation.An external visual inspection was performed on the cycler with no physical damage noted.There were visual indications of fluid found within the cassette compartment.There were no visual indications of particulates found within the cassette compartment.The simulated treatment post-accelerated stress test 2-hour 15-minute 8500 ml test passed.No fluid leaks were found in the test cassette during treatment.The cycler underwent and passed a patient sensor check.An internal visual inspection identified evidence of dried fluid within the recess of the bottom cover adjacent to the pump.The cause of the dried fluid could not be determined.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, there were no malfunctions that could have caused or contributed to the reported event.The cycler performed as designed and an associated cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY PDX CYCLER ASSY (NON-VALUATED)
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
MDR Report Key10088292
Report Number2937457-2020-00930
Device Sequence Number11494163
Product Code FKX
UDI-Device Identifier00840861100989
UDI-Public00840861100989
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K141145
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number180237
Device Catalogue NumberRTLR180237
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Device AgeMO
Initial Date Received by Manufacturer 05/15/2020
Supplement Date Received by Manufacturer06/26/2020
Initial Report FDA Received Date05/26/2020
Supplement Report FDA Received Date07/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID.; LIBERTY PDX, INTEGRATED SET.
Patient Age25 YR
Patient Weight72
-
-