| Model Number |
CD3257-40 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented remotely via merlin.Net.Upon review, it was found that the patient's implantable cardioverter defibrillator was exhibiting far r-wave over-sensing that led to inappropriate automatic mode switching.No intervention was performed.The patient experienced no adverse consequences.
|
| |
|