Additional information: d10, h3, h6 the reported event of unable to remove the right atrial lead and right ventricular lead from the device header was not confirmed.The header was broken apart with only pieces of it returned.No foreign material or anomalies were noted in the pieces that would cause the inability to remove the leads from the header.No further analysis could be performed due to the damage to the header and an insufficient portion of the header was returned from the field to make an evaluation of the problem.Furthermore, the device battery voltage was above elective replacement indicator (eri) level.
|