• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC TRANSDUCER X7-2T; ULTRASOUND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC TRANSDUCER X7-2T; ULTRASOUND Back to Search Results
Model Number 989605414121
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event or Problem Description
A customer reported encountering an articulation issue with their x7-2t model transducer during a surgical procedure.There was no injury associated with this event.
 
Additional Manufacturer Narrative
Evaluation of the transducer confirmed the articulation issue as described by the customer.Functional testing of the device noted failure of the image and articulation tests.Visual inspection noted damage to the i-tube.Destructive testing identified an insufficient solder joint allowed the steering cable to be separated from the rack actuator which resulted in the loss of articulation.Retraining of the employees responsible for the solder joints during the manufacturing process is being pursued to eliminate similar recurrences.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSDUCER X7-2T
Common Device Name
ULTRASOUND
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
MDR Report Key10089474
Report Number3019216-2020-00029
Device Sequence Number19787533
Product Code ITX
UDI-Device Identifier00884838061668
UDI-Public(01)00884838061668
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K030455
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number989605414121
Device Lot NumberB1GY4J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/15/2020
Supplement Date Received by Manufacturer05/15/2020
Initial Report FDA Received Date05/26/2020
Supplement Report FDA Received Date08/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-