• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PRSTHSS,HP,SM-CNSTRND, MTL/CRMC/PLYMR,CEMENTEDORNON-POROUS,UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PRSTHSS,HP,SM-CNSTRND, MTL/CRMC/PLYMR,CEMENTEDORNON-POROUS,UNCEMENTED Back to Search Results
Catalog Number 71365705
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/27/2020
Event Type  malfunction  
Event Description
It was reported that during inspection it was found that the handle is broken off.No case involved.No injury reported.
 
Manufacturer Narrative
The device, intended for use treatment was not returned for evaluation, reporting event could not be confirmed.Therefore, product analysis could not be performed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY INSERTER POSTER HARD
Type of Device
PRSTHSS,HP,SM-CNSTRND, MTL/CRMC/PLYMR,CEMENTEDORNON-POROUS,UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10089820
MDR Text Key192143817
Report Number1020279-2020-01864
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010555663
UDI-Public3596010555663
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/16/2020
Device Catalogue Number71365705
Device Lot Number08KM00707
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date04/27/2020
Date Manufacturer Received08/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-