There was no patient involvement.
Livanova (b)(4) manufactures the s5 gas blender system.
The incident occurred in (b)(6).
The affected device was requested back to the manufacturer site for investigation.
Results confirmed the reported issue and functional tests revealed an error code outbreak associated to the reported discrepancy between the actual and set values of o2 output.
The issue was traced back to a defective o2 honeywell sensor.
This part will be replaced as well as three (3) cables harness for air/o2/co2 which will be replaced as precaution.
|