• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLGER ULLRICH YUNO 2 EU WITH AUTODRIVE; TABLE, OPERATING-ROOM, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLGER ULLRICH YUNO 2 EU WITH AUTODRIVE; TABLE, OPERATING-ROOM, AC-POWERED Back to Search Results
Model Number 143302B0
Device Problems Device Difficult to Setup or Prepare (1487); Insufficient Information (3190)
Patient Problem Injury (2348)
Event Date 04/22/2020
Event Type  Injury  
Manufacturer Narrative
The device is manufactured by maquet gmbh.Maquet medical systems is the importer of this device.A follow-up importer medical device report will be submitted once the results from the manufacturer are available.
 
Event Description
An accessory called "traction bar with ball joint" (1007.40a0) can be mounted and detached from the operating table.While mounting this accessory a staff member injured the tip of a finger.A surgery was required to repair the tip of the finger.Manufacturer reference # (b)(4).
 
Event Description
Manufacturer reference # 321645.
 
Manufacturer Narrative
The potentially affected products (100740a0 - traction bar with ball joint) were investigated by a getinge-maquet service engineer.There are two products 100740a0 in the clinic.Their serial numbers are (b)(6).The user could not tell, which of these was involved in the incident.The traction device with serial number 130 was not fully releasing.This makes it more difficult to attache the device to the talble.The traction device with serial number 46 could be attached and detachted from/to the table easily.In the instructions for use (ifu) the user is warned as follows concerning the use of defective devices: "warning! risk of injury! faulty or defective products may result in injuries.Before use, check the proper working order and fully functional state of the product.Stop using faulty or defective products and inform the getinge representative." the injured user could not remember where his hand was, when the incident occurred.The risk of crashing and shearing hazards is inherent due to the size and weight of the product.In the ifu for 100740a0 the user is warned concerning these risk as follows: "warning! risk of injury! whenever the product is mounted and adjusted, there is a danger of pinching and shearing to the staff, patient and accessories.Always ensure that no one can be subjected to pinching or shearing action or injured in any other way and that the accessories do not collide with any nearby objects." the complaints database was reviewed concerning similar issues with 100740a0 or the yuno ii tables (143302xx).No similar issues were reported to us.Getinge-maquet gmbh provides product failure investigation, analysis and resolution for the device described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YUNO 2 EU WITH AUTODRIVE
Type of Device
TABLE, OPERATING-ROOM, AC-POWERED
Manufacturer (Section D)
HOLGER ULLRICH
maquet gmbh
kehler strasse 31,
rastatt 76437
GM  76437
MDR Report Key10091555
MDR Text Key192209690
Report Number3013876692-2020-00025
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial,Followup
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number143302B0
Device Catalogue Number143302B0
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/17/2020
Distributor Facility Aware Date10/27/2020
Event Location Hospital
Date Report to Manufacturer11/17/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/27/2020
Supplement Dates Manufacturer Received10/27/2020
Supplement Dates FDA Received11/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
100740A0 TRACTION BAR WITH BALL JOINT; 100740A0 TRACTION BAR WITH BALL JOINT
Patient Outcome(s) Required Intervention;
-
-