| Model Number |
459010 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per the manufacturer's evaluation: the forceps was disassembled and the slider inside was found broken at the rivet eye.The broken surface shows a fresh ductile break; a part of the slide is missing.The cutting area shows at least on indication for a cut trial of something solid--the cutting edges are not sharp anymore.The ductile appearance indicates a break by overload.During the operation in which the device failed, or in a previous one, there has been a try to cut something rigid--the cutting edge is blunt on several spots.The breakage was most likely due to mechanical overload by trying to cut hard tissue (bones); it is for cartilage only.
|
| |
|
Event or Problem Description
|
|
As per report received from the factory in (b)(6) regarding an event in (b)(6): during a paranasal sinuses procedure, the forceps lever got blocked all of a sudden.A small metal particle came loose.It was not possible to open or close the forceps anymore.The particle was retrieved.
|
| |
|