Model Number 459010
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Per the manufacturer's evaluation: the forceps was disassembled and the slider inside was found broken at the rivet eye.The broken surface shows a fresh ductile break; a part of the slide is missing.The cutting area shows at least on indication for a cut trial of something solid--the cutting edges are not sharp anymore.The ductile appearance indicates a break by overload.During the operation in which the device failed, or in a previous one, there has been a try to cut something rigid--the cutting edge is blunt on several spots.The breakage was most likely due to mechanical overload by trying to cut hard tissue (bones); it is for cartilage only.
 
Event or Problem Description
As per report received from the factory in (b)(6) regarding an event in (b)(6): during a paranasal sinuses procedure, the forceps lever got blocked all of a sudden.A small metal particle came loose.It was not possible to open or close the forceps anymore.The particle was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAMMBERGER RHINOFORCE II ANTRUM PUNCH
Common Device Name
ANTRUM PUNCH
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
MDR Report Key10092940
Report Number9610617-2020-00060
Device Sequence Number11598770
Product Code LRC
UDI-Device Identifier04048551164532
UDI-Public04048551164532
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number459010
Device Catalogue Number459010
Device Lot NumberTT07
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2020
Initial Report FDA Received Date05/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-