| Model Number |
657312S |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device discarded.
|
| |
|
Event or Problem Description
|
|
It was reported: the locking screw did not lock to the plate.Another new screw was opened and used without problem.
|
| |
|
Event or Problem Description
|
|
It was reported: the locking screw did not lock to the plate.Another new screw was opened and used without problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that a locking screw variax2 t10, full thread, 3.5mm / l12mm was alleged of "implant - insertion impaired" could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|