• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ5 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ5 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1570-12-110
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cellulitis (1768); Fever (1858); Unspecified Infection (1930); Pain (1994); No Code Available (3191)
Date of Event 06/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
After review of the medical records the patient was revised to address pain, cellulitis, fever, chills, noted abscess in the ct scan and periprosthetic joint infection.Operative note reported biofilm formation in the head.Doi: (b)(6) 2019; dor: (b)(6) 2019: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(6).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ5 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10094928
Report Number1818910-2020-12701
Device Sequence Number3407557
Product Code LPH
UDI-Device Identifier10603295060161
UDI-Public10603295060161
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K011489
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Model Number1570-12-110
Device Catalogue Number157012110
Device Lot NumberBR8G91000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/12/2020
Supplement Date Received by Manufacturer07/09/2020
Initial Report FDA Received Date05/28/2020
Supplement Report FDA Received Date07/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SUMMIT DUOFIX TAP SZ5 HI OFF; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-