• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HANDLE; HIP INSTRUMENTS : HANDLES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HANDLE; HIP INSTRUMENTS : HANDLES Back to Search Results
Model Number 2440-00-510
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the pinnacle quickset acetabular grater handle (d244000510) was determined to have collected uncleanable bioburnden or at least insufficient confidence to determine it clean and safe for resterilization.They have been asked to remove all of this style grater handles from this facility (qty 10).
 
Manufacturer Narrative
Product complaint #
=
> (b)(4).Investigation summary
=
> the instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HANDLE
Type of Device
HIP INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10094964
MDR Text Key192393274
Report Number1818910-2020-12708
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295123798
UDI-Public10603295123798
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2440-00-510
Device Catalogue Number244000510
Device Lot NumberA0110
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/20/2020
Initial Date FDA Received05/28/2020
Supplement Dates Manufacturer Received06/29/2020
Supplement Dates FDA Received07/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-