It was reported that the baby was not heating on the arctic sun device.
Last night pad was soiled so they changed it.
Since then they got an alert 02 on the device and the baby was below target temperature (37c).
Ms&s asked what the flow number were and the nurse stated 1.
1-1.
5lpm.
Ms&s told that the flow was acceptable for neonate pads.
Complainant documented that patient temperature twice as 33c and the water was 29c for the first and as 36c and the water was 34c for the second.
Ms&s had the nurse check the graph and the patient temperature with another source.
The nurse stated the baby was paralyzed, with no warmers on, or infections.
Then nurse stated there was surface counter-warmers on and that the vent circuit was set to 37c.
Ms&s asked to turn off the counter-warmers.
The nurse could not turn down the vent circuit.
Ms&s asked to call in an hour if still concern for patient not being at goal.
Per follow up, the biomed could not locate the device.
|