• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC EQUIPMENT RACK 9734061 I7 220V; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC EQUIPMENT RACK 9734061 I7 220V; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9734061
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Onsite functional and visual examination was performed by a manufacturer representative.The system passed a system checkout and was determined to be operational.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the facility was unable to transfer patient data through the network.The site reported that they lost some patient data.
 
Additional Manufacturer Narrative
Continuation of concomitant medical products' information references the main component of the system.Other relevant device(s) are: product id: 9 733763, version #: 2.2.7.The software analysis was completed and no logs or exams were available.There was insufficient information to determine whether a software anomaly contributed to the reported behavior.  if information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Correction h3: a medtronic representative went to the site to test the equipment.The reported issue was confirmed.It was determined some patient exams could not be downloaded.The issue was not resolved at the time of system service completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received from a manufacturer representative indicated cleaning the patient record database and upgrading the cranial software resolved the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUIPMENT RACK 9734061 I7 220V
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key10097094
Report Number1723170-2020-01539
Device Sequence Number11609670
Product Code HAW
UDI-Device Identifier00613994667458
UDI-Public00613994667458
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9734061
Device Catalogue Number9734061
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2020
Supplement Date Received by Manufacturer07/23/2020
08/03/2020
Initial Report FDA Received Date05/28/2020
Supplement Report FDA Received Date07/31/2020
08/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-