| Model Number |
1012-04-030 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Not Applicable (3189); No Code Available (3191)
|
| Date of Event |
05/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon broached up to a 3 broach and implanted a size 3 thrilock std.It sat proud and after rebroaching it still sat proud.He then removed the stem and broached up to a 4 then implanted a new size 3.This delayed the case about 5-10 minutes but was not detrimental to the outcome.I will wash and return the implant in question.Doe: (b)(6) 2020 affected site: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4) investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.H10 additional narrative: added: h6 (patient and device codes) h6 patient code: no code available (3191) used to capture the prolonged surgery and insufficient information.H6 device code: no code available (3191) used to capture the functional: trial does not replicate the implant.
|
| |
|