• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L. P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 38, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L. P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 38, E-PLUS Back to Search Results
Model Number 521-07-238
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to failed rotator cuff.The previous surgery and the surgery detailed in this event occurred almost 1 year apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was no delay in surgery and another suitable device was available.The revision surgery was completed as intended and without incident.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to failed rotator cuff.There were no findings during this evaluation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - due to patient revised from anatomic to reverse due to failed rotator cuff.
 
Additional Manufacturer Narrative
D10: adding concomitant part 600-15-100.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE ANATOMIC SHOULDER
Common Device Name
ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 38, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd
austin, tx
Manufacturer (Section G)
ENCORE MEDICAL L. P.
9800 metric blvd
austin, tx
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 
MDR Report Key10097827
Report Number1644408-2020-00471
Device Sequence Number3407469
Product Code KWS
UDI-Device Identifier00190446178354
UDI-Public(01)00190446178354
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2023
Device Model Number521-07-238
Device Catalogue Number521-07-238
Device Lot Number890U1043
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2020
Supplement Date Received by Manufacturer04/30/2020
Initial Report FDA Received Date05/28/2020
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-07-000, LOT 888U1300; 520-10-000, LOT 851U1106; 520-38-214, LOT 946U1003; 600-15-100 COBALT G-HV BONE CEMENT 40GM
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexFemale
-
-