The reported event that cannulated screw, partially threaded asnis iii ø4.0x50mm tl17mm was alleged of ¿implant - deformation intra-operative¿ could be confirmed, since the product was returned and was found to be broken.The device inspection revealed the following: the received screws head shows stripped at the head position and further no any damage was found.The edges of the hexagon deformed due to improper insertion of the screw driver.The surface of the head shows that there was a misalignment during screw insertion.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The ifu clearly states that ¿ensure that you are familiar with the intended uses, indications/contraindications, compatibility and correct handling of the implant, which are described in the operative technique manual for the product system.' the implant extraction set states that ¿to avoid damaging the screw, make sure the screwdriver is in line with the screw axis and fully inserted.Never use a worn or damaged screwdriver to remove screws.It is recommended that the solid screwdriver be used for screw removal.Strong bone formation around the implant is possible in the pediatric cases using partially threaded screws.This may lead to difficult implant removal with an increased risk of screw head breakage or stripping of screw hexagonal head.¿ based on investigation, the root cause was attributed to a user related issue.The failure was caused by wrong positioning / misalignment between implants leads to strip -off.If any further information is provided, the complaint report will be updated.
|