ARTHROCARE CORP. AMBIENT HIPVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 72290004
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/08/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during hip surgery the ambient hipvac presented an error e6 open t/c.The procedure was completed without delay using a s+n backup device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found a related failure.Visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Potential factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: (1) previous use (2) disconnecting the wand from the controller after power has been turned on.(3) an issue with one of the concomitant devices in use at the time of this complaint event.There are no indications to suggest the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBIENT HIPVAC 50 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10100031
Report Number3006524618-2020-00300
Device Sequence Number11433122
Product Code GEI
UDI-Device Identifier00885556614686
UDI-Public885556614686
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K161481
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2022
Device Catalogue Number72290004
Device Lot Number2037541
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/08/2020
Supplement Date Received by Manufacturer07/08/2020
Initial Report FDA Received Date05/29/2020
Supplement Report FDA Received Date07/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-