| Model Number |
RONYX20015UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
05/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During a procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a severely tortuous, severely calcified lesion located in the mid obtuse marginal (om).The device was inspected with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that stent dislodgement occurred during delivery to/at a lesion.It was stated that an attempt was made to recapture with a wire but the wire could not cross and the stent could not be retrieved.The stent remains in the patients om.The patient was reported to be alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|