| Model Number |
60726807 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A bayer service representative responded to the site and replaced the pivot knuckle assembly which restored the injection system to normal operation.Bayer product analysis examined and analyzed the returned pivot knuckle assembly.Visual examination confirmed a sheared bushing screw within the pivot knuckle assembly.The sheared bushing screw allowed the pivot knuckle to release and the injector head to disengage.
|
| |
|
Event or Problem Description
|
|
The customer reported the following: as the technologist was maneuvering the injector into position for a patient exam, the injector head disengaged from the mounting structure.The injector head was caught by the technologist which prevented it from falling to the floor.No injury or adverse event was reported.
|
| |
|
Search Alerts/Recalls
|