| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed to discharge.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive debug and final testing without duplicating the report.The device was recertified and returned to the customer.The electrode pads used at the time of the event were not returned for evaluation.Without the suspect pads returned for evaluation, we cannot confirm the customer's report.No trend is associated with reports of this type.
|
| |
|
Search Alerts/Recalls
|