| Model Number |
PED-500-14 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Codes |
Death (1802); Intracranial Hemorrhage (1891)
|
| Date of Event |
05/08/2020
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a pipeline device that failed to open at the proximal end.The patient was being treated for an aneurysm of the left ophthalmic artery.Vessel tortuosity was moderate.It was reported that the pipeline device was implanted and the pushwire connection was lost.It was then determined that the proximal end of the device had not opened.The surgeon was unable to retrieve or open the pipeline device.In the days following the procedure, the patient developed complications and died.Ancillary devices: navien058 guide catheter, phenom027 microcatheter, asahi chikai10 guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Codes updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received from a manufacturing representative (rep) indicated the devices were prepared per the instructions for use (ifu), and it was clarified that "wire access was lost" meant the pipeline failure to open was observed after implanting an detaching the device.Multiple attempts were made to retrieve the device using snares and attempting access from the other side, but nothing worked.The pipeline had not been placed in a vessel bend, there was no resistance or friction more than expected, and no damage observed to the pushwire or catheter.The patient experienced entercranial bleeding prior to the death.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|