MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-500-14
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Codes Death (1802); Intracranial Hemorrhage (1891)
Date of Event 05/08/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a pipeline device that failed to open at the proximal end.The patient was being treated for an aneurysm of the left ophthalmic artery.Vessel tortuosity was moderate.It was reported that the pipeline device was implanted and the pushwire connection was lost.It was then determined that the proximal end of the device had not opened.The surgeon was unable to retrieve or open the pipeline device.In the days following the procedure, the patient developed complications and died.Ancillary devices: navien058 guide catheter, phenom027 microcatheter, asahi chikai10 guidewire.
 
Additional Manufacturer Narrative
H6.Codes updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received from a manufacturing representative (rep) indicated the devices were prepared per the instructions for use (ifu), and it was clarified that "wire access was lost" meant the pipeline failure to open was observed after implanting an detaching the device.Multiple attempts were made to retrieve the device using snares and attempting access from the other side, but nothing worked.The pipeline had not been placed in a vessel bend, there was no resistance or friction more than expected, and no damage observed to the pushwire or catheter.The patient experienced entercranial bleeding prior to the death.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10102599
Report Number2029214-2020-00509
Device Sequence Number16744220
Product Code OUT
UDI-Device Identifier00847536019638
UDI-Public00847536019638
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/16/2023
Device Model NumberPED-500-14
Device Catalogue NumberPED-500-14
Device Lot NumberA993675
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2020
Supplement Date Received by Manufacturer05/29/2020
05/29/2020
Initial Report FDA Received Date05/29/2020
Supplement Report FDA Received Date06/10/2020
06/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-