|
|
| Catalog Number |
OM-8850 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a hip capsule closure when using a speedstitches, when deploying the needles to pass the sutures or tape, the needles tip would bend and in some cases it broke off at the tip.The procedure was successfully completed with a backup device, no significant delay reported.No further complications reported.All available information has been disclosed.If additional information should be become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6 the device reported, used in treatment, was not returned for evaluation.Although the device is not available, a relationship between the product and reported incident is possibly established and the complaint is more than likely confirmed as the event reported matches an issue identified during manufacturing of the device.Clinical/medical evaluation was completed and concluded that based on the information provided, the root cause of the reported events is the needles being made of incorrect material vs the design 304 spring wire stainless steel.The patient impact beyond the reported retained tips/pieces in the capsule could not be determined; however, currently unable to rule out iatrogenic effects from any possible retained needle tips.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the device reported, used in treatment, was not returned for evaluation.Although the device is not available, a relationship between the product and reported incident is possibly established and the complaint is more than likely confirmed as the event reported matches an issue identified during manufacturing of the device.A review of manufacturing records for the reported pn number (b)(4) found non-conformances or anomalies during manufacturing process related to the reported event.A complaint history review has found related failures.Review of the product ifu found adequate warnings and precautions to prevent damage to the device during use.Risk management documents were reviewed finding no additional risks that require to be added to the reference document.A review of the product/print specifications found material discrepancy used during manufacturing.Clinical/medical evaluation was completed and concluded that based on the information provided, the root cause of the reported events is the needles being made of incorrect material vs the design 304 spring wire stainless steel.The patient impact beyond the reported retained tips/pieces in the capsule could not be determined; however, currently unable to rule out iatrogenic effects from any possible retained needle tips.It was communicated that the patients will return for imaging at 3 months for routine follow-up.No further medical assessment can be rendered at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.The factor related to the design or manufacture of the device that lead to the failure reported include, but are not limited to: (1) supplier related issue - failed to provide material that meet specification.A corrective action was initiated to mitigate recurrence of the reported event and a field action was initiated to recall the product.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|
Search Alerts/Recalls
|
|
|