| Model Number |
0279250101 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the ceramic broke off of the device during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the ceramic insulation on the tip was broken off inside the foot joint.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting removing the probe, a probe inserted into an obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.The probe used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the ceramic broke off of the device during the procedure.All pieces were retrieved.
|
| |
|