• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INTROCAN SAFETY; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INTROCAN SAFETY; CATHETER,INTRAVASCULAR,THE Back to Search Results
Model Number 4252535-02
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 05/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: it was reported that during cleanup after completing an iv insertion, the nurse was punctured on the right index finger.The safety clip was not fully covering the tip of the needle.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Based on the data from our investigation, we are unable to determine the root cause of the reported incident.Review of the discrepancy management system (dsms) database performed for the reported lot number revealed no abnormalities or non-conformances noted during in process or final product inspection.We will maintain this report for further reference and continue to monitor other reports for similar occurrences.If additional pertinent information becomes available a follow-up report will be filed.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Based on the data from the investigation we are unable to determine the root cause of the reported incident.Although the exact root cause could not be determined, user should refer to ifu which states, "never pull the catheter back through the epidural needle due to the possibility of shearing or kinking the catheter.When removing the catheter, excessive force should never be applied.If the catheter becomes difficult to withdraw, consult standard textbooks for specific techniques.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.  we will maintain this report for further references and continue to monitor other reports for similar occurrences.  if any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown, pa
MDR Report Key10105962
Report Number2523676-2020-00148
Device Sequence Number3017515
Product Code FOZ
UDI-Device Identifier04046963165925
UDI-Public04046963165925
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number4252535-02
Device Catalogue Number4252535-02
Device Lot Number0061670100
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/06/2020
Supplement Date Received by Manufacturer05/06/2020
05/06/2020
Initial Report FDA Received Date06/01/2020
Supplement Report FDA Received Date06/17/2020
06/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-