• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC HDO; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC HDO; INTRAOCULAR LENS Back to Search Results
Model Number SOFTECHDO+19.5
Medical Device Problem Code Device Contamination with Body Fluid (2317)
Health Effect - Clinical Code Deposits (1809)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The microscopic examination of the lens indicated that the opacity was superficial, stained for calcium and did not extend through the matrix of the lens.According to neuhann et al, non-lens factors are the likely cause.It would appear that the patient pre-existing condition is responsible for the event.Lenstec can confirm that the lens, its design and the manufacturing process are not at fault due to the extensive testing that we have performed on this model.Furthermore, there have never been any confirmed lens-related cases of clouding, discoloration or opacification for our hema lenses.Lenstec also confirms that all procedures in the manufacturing and packaging of the lens were conducted correctly.
 
Event or Problem Description
Lenstec received a call stating 'cloudy lens explanted and sent for outside lab analysis.'.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HDO
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
MDR Report Key10106464
Report Number9613160-2020-00054
Device Sequence Number3297915
Product Code HQL
UDI-Device Identifier00844369037334
UDI-Public00844369037334
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/11/2019
Device Model NumberSOFTECHDO+19.5
Device Catalogue NumberSOFTEC HDO
Device Lot Number142169
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2017
Initial Report FDA Received Date06/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
-
-