• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD RIGHT SIZE 9; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD RIGHT SIZE 9; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); Scar Tissue (2060); Limited Mobility Of The Implanted Joint (2671)
Date of Event 05/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2020-00145, 3007963827-2020-00146.Concomitant medical products: tibia cemented 5 degree stemmed right size g, item# 42532007902, lot# 64403058.Articular surface fixed bearing cruciate retaining (cr) right 12 mm height, item# 42521000612, lot# 64401203.Report source - (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent manipulation under anesthesia due to stiffness approximately two months and one-week post implantation.There is no additional information at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, g4, g7, h1, h2, h3, h6, and h10: according to tabers medical dictionary ¿ manipulation of a joint is a mobilization technique, sometimes involving a rapid thrust or stretching of a joint, with or without anesthesia.With anesthesia is (mua).Per clinical orthropaedics and related research (how to treat the stiff total knee arthroplasty: a systematic review), mua is used to treat and resolve arthrofibrosis (scar tissue).This procedure is performed to increase articular motion and reduce chronic pain from arthrofibrosis.The indication for manipulation under anesthesia is procedure related due to adhesions and subsequent stiffness.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD RIGHT SIZE 9
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key10106540
Report Number3007963827-2020-00144
Device Sequence Number2420572
Product Code MBH
UDI-Device Identifier00889024230408
UDI-Public(01)00889024230408
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502606602
Device Lot Number64595673
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/19/2020
Supplement Date Received by Manufacturer09/14/2020
Initial Report FDA Received Date06/01/2020
Supplement Report FDA Received Date09/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
Patient Weight87
-
-