| Model Number |
329416 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of bd ultra fine¿ insulin syringes experienced foreign matter contamination and difficult plunger movement.Product defects were noted during use.The following information was provided by the initial reporter: material no: 329416, batch no: 9028862.Consumer reported finding two syringes with something black on the inside of the barrel.The syringes have not been used.Stated, it looks like something melted on inside, black in color.Stated, the plunger was hard to move.Date of event: 2020 (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of bd ultra fine¿ insulin syringes experienced foreign matter contamination and difficult plunger movement.Product defects were noted during use.The following information was provided by the initial reporter: material no: 329416 batch no: 9028862.Consumer reported finding two syringes with something black on the inside of the barrel.The syringes have not been used.Stated, it looks like something melted on inside, black in color.Stated, the plunger was hard to move date of event: (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: no samples (including photos) were returned as of 19 june 2020 therefore the complaint could not be confirmed and the root cause is undetermined.A review of the device history record was completed for batch# 9028862.All inspections were performed per the applicable operations qc specifications.There was one (1) notification [(b)(4)] noted for smeared stoppers.There were two (2) notifications [(b)(4)] noted for out of spec breakout/ sustaining.There was one (1) notification [(b)(4)] noted for grease on barrel.
|
| |
|
Search Alerts/Recalls
|