• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE SYRINGE 0.3ML 30GA 8MM 10BAG 500 WAL; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE SYRINGE 0.3ML 30GA 8MM 10BAG 500 WAL; PISTON SYRINGE Back to Search Results
Model Number 328511
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 6040778.Medical device expiration date: na.H.4.Device manufacture date: 2016-03-28.Medical device lot #: 6172620.Medical device expiration date: na.Device manufacture date: 2016-08-09.Investigation summary: level a investigation: complaint evaluation / complaint history check for the event(s) that occurred.Severity: s_1__; occurrence: a complaint history check was performed and this is the 1st related complaint for label information missing (expiration date) on lot(s) # 6040778 and 6172620.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch # 6040778.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.This catalog was obsoleted and did not have an expiration date.A review of the device history record was completed for batch # 6172620.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.This catalog was obsoleted and did not have an expiration date.Investigation conclusion: as no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
 
Event or Problem Description
It was reported that syringe 0.3ml 30ga 8mm 10bag 500 wal was missing label information.This was discovered before use.The following information was provided by the initial reporter: material no.328511, batch no.6040778, 6172620.It was reported that expiration date is missing from box.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.3ML 30GA 8MM 10BAG 500 WAL
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key10107388
Report Number1920898-2020-00602
Device Sequence Number2152655
Product Code FMF
UDI-Device Identifier00681131311724
UDI-Public00681131311724
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number328511
Device Catalogue Number328511
Device Lot NumberSEE H.10
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2020
Initial Report FDA Received Date06/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-