Model Number MMT-723RNAS
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had no delivery alarm.Customer stated they had tried all remedies.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump wil be returned for analysis.
 
Additional Manufacturer Narrative
On (b)(6)2020 the customer returned pump for an alleged no excessive no delivery/occlusion alarm noted.Device passed the displacement test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.Device history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.The following were noted during visual inspection: cracked case at the display window corners, cracked lcd window, cracked belt clip slot, stained end cap sticker, stained address/serial number label and broken reservoir tube lip.The pump was received with 50% of reservoir tube lip broken off, however the test infusion set and reservoir did lock properly into place.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.Device passed require testing.No excessive no delivery/occlusion alarm noted.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key10107574
Report Number2032227-2020-125093
Device Sequence Number19684862
Product Code OYC
UDI-Device Identifier00643169476677
UDI-Public(01)00643169476677
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAS
Device Catalogue NumberMMT-723RNAS
Device Lot NumberA0723RNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2020
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight260 KG
-
-