Model Number PCU050300130
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use a pacific xtreme pta balloon catheter during procedure.There was no damage to device packaging.There was no issues noted when removing device from hoop/tray.The device was inspected with no issues noted.Balloon inflation difficulty occurred.It was reported that during balloon inflation, balloon wrap occurred and the balloon did not inflate in center.The vessel was not pre dilated.The device did not pass through previously deployed stent.There was no resistance encountered when advancing device.There was no patient injury reported.
 
Additional Manufacturer Narrative
Additional information: a dog -bone/ twist was visible within the balloon in the vessel.A re wrap tool was not used.The procedure was completed using a new balloon.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: event date provided.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACIFIC XTREME
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key10107875
Report Number9612164-2020-02039
Device Sequence Number3407714
Product Code LIT
UDI-Device Identifier08033477076102
UDI-Public08033477076102
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K103464
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/26/2022
Device Model NumberPCU050300130
Device Catalogue NumberPCU050300130
Device Lot Number219018139
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer06/11/2020
06/16/2020
09/23/2020
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date06/16/2020
06/18/2020
09/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-