| Model Number |
MMT-523RNAP |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that that the insulin pump alarmed compromised force sensor system and drive support cap was sticking out.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received unable to perform all functional testing including the displacement, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test or verify a33 alarm due to completely broken reservoir tube lip.Device received with lose drive support disk and broken reservoir tube lip.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.
|
| |
|