MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT ONE; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT ONE; BLOOD GLUCOSE METER Back to Search Results
Model Number 7818
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the covid-19 pandemic, the investigation for this event will be delayed.We will however, send the follow-up report as soon as our laboratory services are able to return to normal.The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that one of his blood glucose readings was not stored in the memory of the contour next one meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
 
Additional Manufacturer Narrative
The customer did not return the device for evaluation.Therefore, the in-house contour next one meter was tested with the in-house contour next test strips using a blood sample, which gave a satisfactory performance.The blood glucose readings obtained during in-house testing were properly stored in the meter's memory.Additionally, a device history record was reviewed for the suspected contour next one meter and no manufacturing anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT ONE
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla, ny
MDR Report Key10108241
Report Number1810909-2020-00280
Device Sequence Number3448441
Product Code NBW
UDI-Device Identifier00301937818011
UDI-Public00301937818011
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 05/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number7818
Device Catalogue Number7818
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/06/2020
Supplement Date Received by Manufacturer12/30/2020
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date01/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-