• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA AG FREESTYLE W TOTE PT-A 1CT EN; PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA AG FREESTYLE W TOTE PT-A 1CT EN; PUMP, BREAST, POWERED Back to Search Results
Model Number 101034712
Health Effect - Clinical Code Inflammation (1932)
Date of Event 05/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service conducted troubleshooting with the customer, including disassembling, inspecting, cleaning and reassembling the kit and tubing set.Following troubleshooting, the customer indicated that the troubleshooting did not resolve the issue.The customer was sent a replacement pump and return of her original pump was requested for testing/evaluation.The customer was contacted by a complaint handler on multiple occasions, including in writing, to get additional information, with no response as of the date of this report.Based on the results of (b)(4), it cannot be definitively concluded that the pump caused or contributed to the customer's mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 0.008% for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history or mastitis." riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.Mastitis requires prompt medical attention for the mother for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On (b)(6) 2020, the customer alleged to medela llc that she was experiencing low suction with her freestyle breast pump, even on the maximum vacuum setting.She additionally alleged that she was developing mastitis, for which she was prescribed antibiotics.
 
Additional Manufacturer Narrative
The device was evaluated on 09/17/2020 with the customer's parts as received and it failed vacuum testing.The vacuum test was then conducted using a lab kit with the customer's pump and it also failed.The customer's report of low suction was confirmed.It was identified in the evaluation that there was damage on the membrane plate on the motor aggregate (refer to attached picture and product evaluation), which confirms the customer's report of low suction; however, ca11-001 found that it cannot be definitively concluded that a low vacuum breast pump causes or contributes to mastitis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE W TOTE PT-A 1CT EN
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA AG
lattichstrasse 4b
baar zug 6341,
MDR Report Key10108390
Report Number1419937-2020-00052
Device Sequence Number14337641
Product Code HGX
UDI-Device Identifier00020451330921
UDI-Public020451330921
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K150499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101034712
Device Catalogue Number101034712
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Initial Date Received by Manufacturer 05/07/2020
Supplement Date Received by Manufacturer09/17/2020
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date10/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-