• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EXPRT REVISION HIP; EXPRT REVISION HIP, STANDARD OFFSET PROXIMAL BODY, 75MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EXPRT REVISION HIP; EXPRT REVISION HIP, STANDARD OFFSET PROXIMAL BODY, 75MM Back to Search Results
Model Number 495-00-075
Health Effect - Clinical Codes Fall (1848); Pain (1994)
Date of Event 05/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported due to stability and pain after fall.The previous surgery and the surgery detailed in this event occurred 1.5 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to stability and pain after fall.There were no findings during this evaluation that indicate that the reported devices were defective.No other information was submitted with the complaint regarding preexisting conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient was unstable and has pain from a fall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT REVISION HIP
Common Device Name
EXPRT REVISION HIP, STANDARD OFFSET PROXIMAL BODY, 75MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
MDR Report Key10108749
Report Number1644408-2020-00425
Device Sequence Number16932933
Product Code KWZ
UDI-Device Identifier00190446153825
UDI-Public(01)00190446153825
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/21/2023
Device Model Number495-00-075
Device Catalogue Number495-00-075
Device Lot Number541U1005
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2020
Initial Report FDA Received Date06/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-364 LOT 869B1104
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-