Model Number 2420-0007 |
Device Problems
Fluid/Blood Leak (1250); Tear, Rip or Hole in Device Packaging (2385); Material Split, Cut or Torn (4008)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/20/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
|
Event Description
|
It was reported from the day surgery unit that just prior to use, the newly primed primary tubing set leaked "normal iv fluid".There was a "minimal delay" caused because a new tubing set had to be prepared.It was observed that the packaging had a little cut on it which the customer thinks is how the tubing set became cut also.There was no patient involvement as this happened prior to use.
|
|
Event Description
|
It was reported from the day surgery unit that just prior to use, the newly primed primary tubing set leaked "normal iv fluid".There was a "minimal delay" caused because a new tubing set had to be prepared.It was observed that the packaging had a little cut on it which the customer thinks is how the tubing set became cut also.There was no patient involvement as this happened prior to use.
|
|
Manufacturer Narrative
|
The customer¿s report that the tubing set leaked was confirmed.Visual inspection observed a cut/slice on the tubing at 9 inches below the distal adapter tubing.The packaging pouch received was also visually inspected, and it was also noted to have slice cut.A leak was observed from the cut tubing.The cause of the leak in the tubing was identified as damage from a slice cut in the packaging pouch.The root cause of this failure was not identified.Device history record for model 2420-0007 lot 20033350 shows that the set was manufactured on 24 march 2020 with a total of (b)(4) units.There were no qn¿s (quality notification) issued during the production build of this lot for the failure mode reported.
|
|
Search Alerts/Recalls
|