• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN OFFSET COUPLER 4MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAIN, CEMNT, PLYMR/MET/PLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN OFFSET COUPLER 4MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAIN, CEMNT, PLYMR/MET/PLYMR Back to Search Results
Model Number 71424225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Limited Mobility Of The Implanted Joint (2671)
Date of Event 05/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an arthroscopy was performed due to poor range of motion.Significant soft tissue was removed.Good range of motion at the end of the surgery.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms reportedly the ¿poor rom¿ secondary to incomplete rehab/severe burns was the root cause of the arthroscopy/revision.The patient impact beyond the reported symptoms and subsequent arthroscopy/revision could not be determined.Although the complaint indicated there was good rom on the table, the actual procedure achieved only ¿minimum results¿ post-operatively per complaint/ response.No further medical assessment can be rendered at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.No further medical assessment is warranted at this time.A review of the risk management file revealed this failure mode was previously identified.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN OFFSET COUPLER 4MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAIN, CEMNT, PLYMR/MET/PLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10110877
Report Number1020279-2020-01940
Device Sequence Number19663386
Product Code JWH
UDI-Device Identifier03596010547132
UDI-Public03596010547132
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K043440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71424225
Device Catalogue Number71424225
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 05/06/2020
Supplement Date Received by Manufacturer06/11/2021
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-