| Model Number |
010001 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a knee arthroscopy, the acufex straight probe broke.It is unknown how the surgery was finished or if there was a surgical delay.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported straight probe, intended for use in treatment, has not been returned for evaluation.Without the reported product a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, ¿during a knee arthroscopy, the acufex straight probe broke¿.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive force applied during use.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|