Model Number 010001
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a knee arthroscopy, the acufex straight probe broke.It is unknown how the surgery was finished or if there was a surgical delay.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported straight probe, intended for use in treatment, has not been returned for evaluation.Without the reported product a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, ¿during a knee arthroscopy, the acufex straight probe broke¿.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive force applied during use.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROBE STRAIGHT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10112701
Report Number1219602-2020-00837
Device Sequence Number2626207
Product Code NBH
UDI-Device Identifier03596010021533
UDI-Public03596010021533
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number010001
Device Catalogue Number010001
Device Lot Number50745618
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/08/2020
Supplement Date Received by Manufacturer07/13/2020
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date07/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-