Model Number CD3265-40
Medical Device Problem Codes Failure to Interrogate (1332); Battery Problem (2885)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.The investigation results will be provided in the final report.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
It was reported that the patient presented in clinic for follow up.Upon interrogation, it was revealed that the implantable cardioverter defibrillator could not be interrogated.Upon review of that last remote transmission, it was revealed that the implantable cardioverter defibrillator received a battery performance alert (bpa) on (b)(6) 2020.Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient was stable post procedure and discharged.
 
Additional Manufacturer Narrative
Communication failure and premature battery depletion were confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA (DDQP+)
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10112771
Report Number2017865-2020-06942
Device Sequence Number14407062
Product Code NIK
UDI-Device Identifier05414734505666
UDI-Public05414734505666
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2014
Device Model NumberCD3265-40
Device Catalogue NumberCD3265-40
Device Lot Number4053926
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/15/2020
Supplement Date Received by Manufacturer07/22/2020
Initial Report FDA Received Date06/02/2020
Supplement Report FDA Received Date08/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0117-2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
-
-