• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Catalog Number 71118200
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/04/2020
Event Type  malfunction  
Event Description
It was reported that while doing instrument inspection it was found that the reamer shaft is broken.No injuries or delays reported.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection showed that the flexible reamer shaft fractured within the shaft via mechanical overload.If sufficient offset bending or torsional forces are applied to the reamer shaft during use that exceeds the strength of the material, a mechanical overload fracture can occur.The device was manufactured in 2017.This device shows significant signs of wear/usage.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10113703
MDR Text Key193584403
Report Number1020279-2020-01960
Device Sequence Number1
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public3596010436528
Combination Product (y/n)N
PMA/PMN Number
K121714
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71118200
Device Lot Number17LM07966
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-