| Model Number |
MMT-551NAP |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The device was not bumped or dropped prior to the alarm and the drive support cap appears protruded.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with a35 alarmed during rewind due to motor encoder signal out of phase.Unable to perform basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test or verify a33 alarm due to a35 alarms.Device received with loose drive support disk, broken battery tube threads and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.
|
| |
|