U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. SL-PLUS AND SLR-PLUS; HEAD, FEMORAL COCR 22MM NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. SL-PLUS AND SLR-PLUS; HEAD, FEMORAL COCR 22MM NEUTRAL Back to Search Results
Model Number 400-01-221
Health Effect - Clinical Code Fall (1848)
Date of Event 05/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported due to dislocation/disassociation after a fall.The previous surgery and the surgery detailed in this event occurred 33 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation/disassociation after a fall.There were no findings during this evaluation that indicate that the reported devices were defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient fell and dislocated inner diameter 22.Head disassociated from 42 outer diameter head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL-PLUS AND SLR-PLUS
Common Device Name
HEAD, FEMORAL COCR 22MM NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
MDR Report Key10114739
Report Number1644408-2020-00430
Device Sequence Number13988903
Product Code JDI
UDI-Device Identifier00888912075626
UDI-Public(01)00888912075626
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/11/2025
Device Model Number400-01-221
Device Catalogue Number400-01-221
Device Lot Number777B1237
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2020
Initial Report FDA Received Date06/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-