BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 26924
Medical Device Problem Codes Entrapment of Device (1212); Premature Activation (1484); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: patient was over 18 years old.
 
Event or Problem Description
It was reported that the stent inadvertently deployed and the delivery system became froze on the wire and fractured.The target lesion was located in the right superficial femoral artery.Popliteal access was used with a 6fr super sheath.Two stents and balloons were used over the amplatz guidewire.An 8x80x130 innova self expanding stent was selected for use.During preparation for the procedure, the stent and delivery systems were appropriately flushed prior to loading and lock was in place.Upon insertion over the guidewire and during forward advancement outside the patient body, it was noted that the stent had partially deployed approximately 2cm.The physician attempted to remove the catheter but it had become stuck to the guidewire and could not be removed without using significant force which led to the inner shaft fracturing.The fragments were removed from the guidewire with extreme force.The procedure was completed successfully using another innova stent and different guidewire.No patient harm or complications were reported.
 
Additional Manufacturer Narrative
A2: age at time of event: patient was over 18 years old.Device evaluated by manufacturer.Returned product consisted of an innova self-expanding stent system.The stent component was not returned.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed multiple kinks on the outer sheath, inner liner, and proximal inner.The proximal inner and inner liner is separated from the clip.Microscopic examination revealed no additional damages.The pull handle and yellow thumbwheel lock is still in the manufactured position.Inspection of the remainder of the device, revealed no other damage or irregularities.
 
Event or Problem Description
It was reported that the stent inadvertently deployed and the delivery system became froze on the wire and fractured.The target lesion was located in the right superficial femoral artery.Popliteal access was used with a 6fr super sheath.Two stents and balloons were used over the amplatz guidewire.An 8x80x130 innova self expanding stent was selected for use.During preparation for the procedure, the stent and delivery systems were appropriately flushed prior to loading and lock was in place.Upon insertion over the guidewire and during forward advancement outside the patient body, it was noted that the stent had partially deployed approximately 2cm.The physician attempted to remove the catheter but it had become stuck to the guidewire and could not be removed without using significant force which led to the inner shaft fracturing.The fragments were removed from the guidewire with extreme force.The procedure was completed successfully using another innova stent and different guidewire.No patient harm or complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10115260
Report Number2134265-2020-07402
Device Sequence Number11203211
Product Code NIP
UDI-Device Identifier08714729874195
UDI-Public08714729874195
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2021
Device Model Number26924
Device Catalogue Number26924
Device Lot Number0022591651
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Initial Date Received by Manufacturer 05/12/2020
Supplement Date Received by Manufacturer07/09/2020
Initial Report FDA Received Date06/03/2020
Supplement Report FDA Received Date07/29/2020
Patient Sequence Number1
-
-