| Model Number |
7005 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.0mm x 8mm quantum maverick balloon catheter was unpacked, however; the balloon was fractured.The procedure was completed with another of the same device.This was not used inside the patient's body.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 79.6cm distal of the strain relief.There was contrast in the inflation lumen and balloon.The balloon was tightly folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.0mm x 8mm quantum maverick balloon catheter was unpacked, however; the balloon was fractured.The procedure was completed with another of the same device.This was not used inside the patient's body.
|
| |
|