• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7005
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/26/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 3.0mm x 8mm quantum maverick balloon catheter was unpacked, however; the balloon was fractured.The procedure was completed with another of the same device.This was not used inside the patient's body.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 79.6cm distal of the strain relief.There was contrast in the inflation lumen and balloon.The balloon was tightly folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that shaft break occurred.A 3.0mm x 8mm quantum maverick balloon catheter was unpacked, however; the balloon was fractured.The procedure was completed with another of the same device.This was not used inside the patient's body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10115638
Report Number2134265-2020-06550
Device Sequence Number16606317
Product Code LOX
UDI-Device Identifier08714729392569
UDI-Public08714729392569
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2022
Device Model Number7005
Device Catalogue Number7005
Device Lot Number0024395604
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Initial Date Received by Manufacturer 05/08/2020
Supplement Date Received by Manufacturer06/25/2020
Initial Report FDA Received Date06/03/2020
Supplement Report FDA Received Date07/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Patient Weight80
-
-