| Model Number |
MMT-551LNAB |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump esc and act button need to press multiple times to work.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to flattened express bolus, esc, act, up and down button dome switch.No button error alarm during testing.No unlocked j2/lcd keypad connector.Device passed self test.
|
| |
|