MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAS
Medical Device Problem Code Insufficient Flow or Under Infusion (2182)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 05/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced hyperglycemia.The customer¿s blood glucose level was more than 400 mg/dl.Customer had been using insulin pump system within 48 hours of reported high blood glucose event.The customer declined troubleshooting.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with all operating currents within specification and passed functional testing including the rewind, a21 error test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.No possible under delivery anomaly noted.Device was downloaded to carelink pro properly and all bolus delivered were found at the carelink report.Device passed the delivery accuracy test at 0.0874 inches.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10118029
Report Number2032227-2020-126253
Device Sequence Number16606408
Product Code OYC
UDI-Device Identifier00643169513860
UDI-Public(01)00643169513860
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA4523LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2020
Initial Date Received by Manufacturer 05/28/2020
Supplement Date Received by Manufacturer06/15/2020
Initial Report FDA Received Date06/04/2020
Supplement Report FDA Received Date06/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN-UNK-RSVR, UNOMED SET
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
Patient Weight123
-
-