| Model Number |
DBP-CART-125MIC145 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
|
| |
|
Event or Problem Description
|
|
Vessel spasm was observed on imaging during a procedure in which a diamondback peripheral orbital atherectomy device was used.The spasm resolved without intervention by the conclusion of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis conclusion: the cartridge was received without the oad handle for analysis.There was no visible damage on the cartridge.When functionally tested, the cartridge operated as intended.At the conclusion of the failure analysis investigation the reported vessel spasm could not be confirmed.The cause of the reported vessel spasm is unknown.Per the opinion of the physician, the spasm may have been caused by a combination of factors and not just by use of the oad.The diamondback 360® peripheral orbital atherectomy system exchangeable series instructions for use manual states that vessel spasm is a possible adverse event which can occur with use of the system.Csi id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b4 (initial report had incorrect date.).
|
| |
|