| Model Number |
24716 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that removal difficulty occurred.A 3.0 x 220 x 90 (4f) sterling balloon was selected for use.During removal, the guide broke making the removal of the balloon impossible.The hospitalization took longer as the patient was given local anesthesia and additional surgical intervention was reformed to remove the guide.
|
| |
|
Event or Problem Description
|
|
It was reported that removal difficulty occurred.A 3.0x220x90 (4f) sterling balloon was selected for use.During removal, the guide broke making the removal of the balloon impossible.The hospitalization took longer as the patient was given local anesthesia and additional surgical intervention was reformed to remove the guide.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.The proximal portion of the device was missing.Approximately 175mm of the distal portion was returned.The inner shaft was separated 30mm from the tip.There was blood in the balloon.The balloon had a circumferential tear 175mm from the tip.The was a longitudinal tear starting at the circumferential tear and extending 39mm distally.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|