• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US ES2 HEX STRAIGHT ROD 5.5 X 45MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US ES2 HEX STRAIGHT ROD 5.5 X 45MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 482807045
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Injury (2348)
Date of Event 01/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: visual inspection indicates that rod dent on 2 of the returned blockers is not uniform on both sides of the blocker implying that the rod might not be placed horizontally in the screw head.Review of x-rays also indicates that one of the rods was placed at an angle and may not be fully contacting the blockers.No visual anomaly was noted on 2 blockers.Device and complaint history records were reviewed and no relevant manufacturing issues or similar complaints were identified.From es2 surgical techniqe: use the counter torque tube as an insertion tube to facilitate the alignment of the blocker within the blades to prevent cross-threading.The counter torque tube can also be used to help direct the rod downward into the screw head if the rod is slightly proud.Final tightening of the blockers is performed using the counter torque tube and the torque wrench.Turn the handle of the torque wrench clockwise to align the two arrows on the torque wrench to achieve the 12nm of torque required to secure the implant construct.Caution: extra caution is advised in the following cases: the rod is not horizontally placed into the screw head, the rod is high in the screw head, and/or an acute convex or concave bend is contoured into the rod.Based on surgical technique and inspection of returned products and x-rays, most likely cause of reported event is rod not fully contacting the blockers in the initial construct / off-loading on blockers.
 
Event or Problem Description
It was reported that 4 mantis redux blockers and 2 rods, implanted in l2/l3, migrated post-operatively.Revision surgery has occurred.This report represents the second of the two rods.
 
Additional Manufacturer Narrative
Visual inspection: visual inspection indicates that rod dent on 2 of the returned blockers is not uniform on both sides of the blocker implying that the rod might not be placed horizontally in the screw head.Review of x-rays also indicates that one of the rods was placed at an angle and may not be fully contacting the blockers.No visual anomaly was noted on 2 blockers.Device and complaint history records were reviewed and no relevant manufacturing issues or similar complaints were identified as all units met stryker specifications.From es2 surgical technique: use the counter torque tube as an insertion tube to facilitate the alignment of the blocker within the blades to prevent cross-threading.The counter torque tube can also be used to help direct the rod downward into the screw head if the rod is slightly proud.Final tightening of the blockers is performed using the counter torque tube and the torque wrench.Turn the handle of the torque wrench clockwise to align the two arrows on the torque wrench to achieve the 12nm of torque required to secure the implant construct.Extra caution is advised in the following cases: the rod is not horizontally placed into the screw head, the rod is high in the screw head, and/or an acute convex or concave bend is contoured into the rod.Based on surgical technique and inspection of returned products and x-rays, most likely cause of reported event is rod not fully contacting the blockers in the initial construct / off-loading on blockers.
 
Event or Problem Description
It was reported that 4 mantis redux blockers and 2 rods, implanted in l2/l3, migrated post-operatively.Revision surgery has occurred.This report represents the second of the two rods.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 HEX STRAIGHT ROD 5.5 X 45MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key10121317
Report Number0009617544-2020-00078
Device Sequence Number1847689
Product Code NKB
UDI-Device Identifier07613327001464
UDI-Public07613327001464
Combination Product (Y/N)N
PMA/510(K) Number
K122845
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number482807045
Device Catalogue Number482807045
Device Lot NumberYDT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2020
Initial Date Received by Manufacturer 05/09/2020
Supplement Date Received by Manufacturer05/28/2020
Initial Report FDA Received Date06/04/2020
Supplement Report FDA Received Date06/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-