| Model Number |
15 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control contacted the customer to request additional information on the patient.The customer informed physio-control that no further patient information is available.Physio-control evaluated the customer's device and was unable to duplicate the reported issue.Physio downloaded the device's electronic records from the event for review and was able to verify the reported issue.Physio performed unrelated repairs.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device would intermittently detect an ecg.In this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
|
| |
|